TBL5063-3 - 50uH 63A Line Impedance Stabilisation Network LISN
The TBL5063-3 is a Line Impedance Stabilization Network for the measurement of conducted emissions within the range of 9 kHz to 30MHz, according to the CISPR 16-1-2 standard. The device is designed for testing 3-phase, AC-powered equipment with supply voltages up to maximum 690V/400V and 63A. Conducted noise can be measured on each phase and on the neutral conductor. The TBL5063-3 has separate RF outputs for each phase and neutral and consequently can be coupled with the TBLM 2 LISN Mate to split the emissions in its common mode and differential mode components . The TBL5063-3 has no integrated high pass filter at the RF outputs, but comes with an external 9 kHz coaxial high pass instead. The LISN is equipped with 63A CEE sockets.
Specifications
- Frequency range: 9 kHz to 30 MHz
- Impedance: 50 Ω ║ (50 µH + 5 Ω)
- Artificial hand: 220 pF + 511 Ω
- Switchable PE: 50 Ω ║ 50 µH
- 250µH pre-filter
- Separate RF outputs for each line and neutral
- Air core inductors
- Line voltage: max 690V/400V / 50 – 60 Hz
- Max. current per line and neutral: 63A @ 23°C each
- 3-phase EUT socket: CEE / IEC60309, 3L+N+PE, 63A, female
- Power connector: CEE / IEC60309, 3L+N+PE, 63A, male
- Measurement connectors: 50 Ω BNC
- External 9 kHz high pass filter, N-male / N-female
- Operating Temperature Range: +5 °C … + 40 °C; 5% to 80% RH
Cables, connectors and calibrators can be found here.
Documentation
TBL5063-3 50µH LISN User Manual